Software Update 1.40
New rinsing function
The "rinsing function" is a new activity you can use on your MASTERsurg. The rinsing function serves to feed coolant and to start-up the illumination on the handpiece. The motor is not activated during this process in order to protect the patient. The illumination is switched on along with this function. The supplied quantity can be controlled in stepless manner using the speed button on the foot control. To optimise the workflow of treatment, the rinsing function can be programmes as a separate work step just like any other activity. However, the function can also be activated manually by pressing the pump button on the foot control for a longer period of time.
Change of transmission ratios that can be set
The transmission ratio of 1:5 for high-speed handpieces has been replaced by a reduction ratio of 16:1. This setting is designed mainly for the common combination of the following KaVo products: INTRA surgery shank 3624 N (0.534.5660) and INTRA reducing head 67 RIC (0.540.8340). Since this contra-angle handpiece is well-suited for inserting transplants as well, it can be used in one-touch calibration.
Change of the torque profile
The torque curve has been optimised to allow an even higher torque to be reached at high speeds. Accordingly, a torque of 2.5 Ncm at 40,000 rpm is now available to the user.
The present software update applies to devices up to production date 17 Nov. 2014, whereas devices manufactured after this date include this update already.